CORRECTION FOR FDX OPERATION MANUAL
REFER PAGE 23: SECTION F3. INTERCHANGING FORCE CELL MODULES, SELECTING THE FORCE CELL MODULE. REPLACE STEPS 1-3 AS FOLLOWS:
Power off the gage. While holding down ON/OFF, press SET and follow by UNITS. Release ON/OFF to access the calibration menu. The current programmed unit and capacity “2 lbf” is displayed. The blinking display indicates the value is the current programmed unit and capacity.
Use the steps below to select the new capacity.
Press ­ to step through available capacities until “ lbf” is displayed.
Press UNITS to step through available units until “ kgf” is displayed ( “kgf” corresponds to the “lbf” unit setting).
IMPORTANT: UNITS MUST BE SET ON “kgf” MODE WHILE ENTERING THE NEW LOAD CELL SENSITIVITY VALUE. If testing requires the “lbf” or “N” mode, reset UNITS only after exiting the set-up mode to avoid corrupting gage calibration.
3a. Press ENTER to save the new capacity and unit settings. The display blinks and displays
the digit “2” value before the decimal point of the mv/v of the previous Force Cell Module. The digit “2” is blinking to indicate that this digit is being set.
4. CONTINUE TO PAGE 24.
WAGNER FORCE TEN
MODELS FDXE & FDXE-R
Force Gage
Operation Manual
IMPORTANT
READ THIS BEFORE USING THE FORCE GAGE
CAUTION
EXCESSIVE OVERLOAD,
OVER 25% ABOVE RATED CAPACITY,
MAY CAUSE PERMANENT DAMAGE TO THE LOAD CELL.
WAGNER INSTRUMENTS
PO BOX 1217
GREENWICH, CT 06836-1217
TEL: 800 – 345 – 4188 FAX: 800 – 443 - 4149
TABLE OF CONTENTS
A. GENERAL INFORMATION
Operation Features
Physical Design Features
B. INTERFACE FEATURES
Keypad Functions
Power On
Power Off
Setting Automatic Power Off Parameter
C. TEST MODES
AVERAGE Mode
T PEAK Mode
C PEAK Mode
D. INTERCHANGING FORCE CELL MODULES………………………………………………………………5
Connecting a new integral FORCE CELL MODULE (FCMX)………………………………………….5
Co