文档介绍:ERRATUM: OIML R 60
Edition 2000 (E)
INTERNATIONAL
MENDATION
Metrological regulation for load cells
THIS ERRATUM ONLY CONCERNS THE ENGLISH VERSION OF R 60 (2000)
ORGANISATION INTERNATIONALE
DE MÉTROLOGIE LÉGALE
ANIZATION
OF LEGAL METROLOGY
Erratum to OIML R 60(E) (Edition 2000) November 2000
NOTE: Since the publication of OIML R 60(E) Edition 2000, a number of points have been corrected. These are listed
below and should be taken into account when using OIML R 60.
Page in Clause Correction
R 60 (2000) or section to be applied
40 Replace the existing text by:
Divide CM by (T2 – T1) and multiply the result by 5 for classes B, C, and D, or
by 2 for class A. This gives the change in v per 5 °C for classes B, C, and D, or
in v per 2 °C for class A.
40 Replace the existing text by:
Multiply the preceding result by [(Dmax – Dmin)/n] /vmin to give the final result in
units of vmin per 5 °C for classes B, C, and D, or units of vmin per 2 °C for class A;
this final result must not exceed pLC.
44 Replace the symbol “<” by the symbol “≤”:
≥
n = nmax – 500 and n = nmax – 1 000 (provided that n 500)
44 Replace the symbol “<” by the symbol “≤”(twice):
≤
vmin v
≤
vmin (Dmax – Dmin) / nmax
47 Annex D Under the section “Information concerning the pattern”, “Power voltage”, second
line, insert a colon (:) after “V”:
or V: ____________ V
50 Summary Change the reference between the two tables from “” to “”:
of the test
50 Summary Change the second table as below:
of the test
n n – 500 n – 1 000
Paragraph max max max
Description
No. Pass Fail Pass Fail Pass Fail
, Check all calculations using values of n at n
, max
and at lower than n
max
D – D
≤ max min
Check that vmin
nmax
Worst case figure for minimum dead load output return error (in mass units) = DR = _____________ see Note 3
56 Table Left column: Change the reference in the bott