文档介绍:Finite Element Simulation of Drill Bit and Bone
Thermal Contact During Drilling
Yuan-Kun Tu Hsun-Heng Tsai
Director, Department of Orthopedics Department of Biomechatronic Engineering
E-Da Hospital and I-Shou University National Pingtung University of Science and Technology
Kaohsiung Hsien, Taiwan Pingtung, Taiwan
******@.tw ******@
Li-Wen Chen, Ching-Chieh Huang and Yung-Chuan Chen* Li-Chiang Lin
Department of Vehicle Engineering Director, Department of Biomedical
National Pingtung University of Science and Technology E-Da Hospital and I-Shou University
Pingtung, Taiwan Kaohsiung Hsien, Taiwan
******@ (*corresponding author) ******@.tw
Abstract— Drilling is an essential part of internal fixation in results explored by Bachus et al. [3] also showed that by the
orthopaedic and trauma surgery. The temperature rise during application of a larger force to the drill, both maximum cortical
bone drilling is an important index to the damage of bone. In this temperatures and their duration above 50 oC might be
study, an elastic-plastic dynamic finite element model is used to effectively reduced. Allan et al. [4] reported that the
simulate the process of a drill bit drilling through the bone.
Various initial temperatures of drill bit are investigated to temperature rise of drilling bone was related to the amount