文档介绍:Prodigy型骨密度仪测量骨密度的变异系数研究
(作者:___________单位: ___________邮编: ___________)
作者:郭郡浩,杨路昕,赵智明,臧旭丽
【摘要】目的分析骨密度仪测量骨密度的变异系数。方法使用双能X线骨密度仪测量标准模块,每日1次,共测量1156次,根据测量结果计算变异系数。结果骨密度仪测量高、中、,%、%、%。结论骨密度分析仪测量骨密度变异系数较低,有较好的稳定性,但其变异系数随着骨密度的降低有增加趋势。
【关键词】骨质疏松;骨密度;双能X线吸收法;变异系数;质量控制
Abstract: Objective To analyze the coefficient of variation (CV) by measurements of bone mineral density with the dual-energy X-ray absorptiometer (DXA, Prodigy). Methods A standard calibration module was measured by daily Quality Assurance (n=1156), and the value of CV was calculated and analyzed according to the measured data. Results The standard deviation of BMD measurement was , in the order of high, middle and low BMD, and the CV was %, % and %, respectively. Conclusion The DXA measurement of BMDs reveals low CVs, which indicates the satisfactory stability of DXA, while the CVs tend to increase in accordance with the decrease of BMDs.
Key words: osteoporosis; bone mineral density; dual-energy X-ray absorptiometry; coefficient of variation; quality assurance
目前,应用双能X线吸收法(DXA)测量骨密度(BMD)被视为诊断骨质疏松症的“金标准”[1]。与其他检查设备一样,DXA骨密度仪也同样存在着测量误差。为了进一步分析骨密度仪测量骨密度的变异系数(coefficient of variation,CV),我们进行了本研究,现将研究结果报告如下。
1 资料和方法
数据资料数据来源于本院骨密度检查室,时间为2004年2月17日至2009年1月27日,共1156次测量结果,均全部完成测量程序。
骨密度分析仪 DXA骨密度分析仪由美国GE公司生产,型号为GE Lunar Prodigy型。
标准模块测量骨密度用的标准模块由