文档介绍:Detection of pulmonary nodules among monochrome
LCDs with different resolutions – an observer
performance study

Jiandong Yin, Qiyong Guo Xianzheng Sha
Department of Radiology Department of Biomedical Engineering
Shengjing Hospital of China Medical University China Medical Engineering
Shenyang, Liaonin Shenyang, Liaoning
jiandongyin@ ******@.


Abstract—Objective To evaluate the influence of monochrome one pare observer performance using the same brand of
LCDs with different resolutions on the detection performance of medical monochrome LCDs with different resolutions for the
pulmonary nodules. Materials and Methods 93 chest images were detection of pulmonary nodules. Full consideration was given
selected, including 55 normal images and 38 positive images. to the human factors in this research. This study was designed
Three high-, three mid-, and three low-experienced radiologists to simulate daily clinical practice, and its conclusion could
interpreted these images using three types of displays. Regarding provide a scientific basis for rational allocation of diagnostic
the presence of nodule, five-point confidence level rating scale display in the radiology department. That meaning is very
was used: definite absence, probable absence, indetermination, important currently for some developing countries because of
probable prese