文档介绍:·674· 检验医学 2021 年 6 月第 36 卷第 6 期 Laboratory Medicine,June 2021,Vol. 36,No. 6
xed serum samples for these 4 items were tested throughout the
study. The results of electronic quality control,liquid quality control and mixed serum samples were analyzed. Results
Each result of electronic quality control passed in 2 phases. The liquid quality control results were within control when
they were assessed adopting Westgard sigma rules in different phases. The coefficient of variation (CV)of the mixed
serum sample results for 4 items were all <1/3 allowable total error (TEa). Conclusions When microfluidic chip
POCT is used to test K,Na,Cl and Glu,under the condition of stable determination personnel and environment,
the same batch of reagents,the use of daily electronic quality control with weekly liquid quality control can ensure the
quality of analysis.
Key words:Microfluidic chip;Point-of-care testing;Internal quality control
随着即时检测(point-of-care testing, 家协会规定对POCT设备必须每月检测1次质控
POCT)在临床科室的广泛应用,其检验结果的 品,对新批号的试剂也必须检测质控品[2]。在瑞
质量控制成为一个迫切需要解决的问题。发达 士,IQC频次根据分析系统的操作繁易程度确
国家POCT发展较早,已建立起完善而规范的管